Upper Y axis and radius holder  - SPN
Upper Y axis and radius holder  - SPN

Upper Y axis and radius holder - SPN


notion image

Required Tools and Components


Components

  • Exchange plate and adapter plate
  • Argon load cells
  • Graphite plate
  • Holder plate
  • (Optional) Suspension
  • Ball/block holder
  • (Optional) Block extension

Screws and Hardware

  • 4 x 6-32 x .250” screws
  • 4 x 10-32 x .438” screws
  • 4x BN610 M4 x 8 screws
  • Imperial Allen keys set
  • Metric Allen keys: 3mm
 

Step-by-Step Installation


Remove the current adapter and holder if present


Every accessories must be removed along with the graphite plate.

The graphite plate will be mounted back in the next part.
notion image
 

Install the adapter plate


  • Position the rectangular plate along the Y axis of the load cell to support the module.
page icon
6-32 x .250” long using 7/64” Allen wrench
 
notion image

Install the Y axis module


  1. Firstly, lose the tightening screw on the right to free the upper plate and have access to each screws.
  1. Move the upper plate to 40mm and re tighten the side screw.
  1. Secure the upper module with 4 x M4 x 8 screws using a 3mm metric Allen key.
  1. Move back the plate to thighten the 2 last screws.
notion image

Install the graphite plate and the holder


  1. Fix the second adapter with four 6-32 x .400” screws using a 3mm metric Allen key.
  1. Secure the graphite plate with four 6-32 x .250” screws long using 7/64” Allen wrench.
  1. Tighten the two captive screws from the suspension using 9/64” Allen key.
Without suspension

  • Mount the graphite plate
  • Fix the holder with 4 x 10-32 x .438” using 5/32” Allen wrench
    • notion image
 
 
 
 
notion image
 

Center the holder to 0


50mm of total stroke length, considering 25mm is the center point.

⚠️
The side screws must be loosen first.
  1. Adjust the micrometer screw to increase or decrease the Y offset manually.
  1. Tighten the side screw.
 
notion image
 
 

Pin/Ball holder preparation


The ball holder along with the provided collet can also accommodate circular pins.

notion image
  1. Loosen the nut to free the ball.
  1. Insert the adjusting pin into the holder, then the ball.
    Provided for standard test: Ball, .250" (1/4") (6.350mm) Dia
    E52100 100Cr6 grade 25 Alloy Steel.
  1. Hold the holder vertically, so the ball is resting on the pin.
    Using a 1/8" Allen key, fasten the screw inside the holder to slightly push the ball.
  1. Once the ball is retracted enough, fasten the nut to secure it.
 
notion image
page icon
For preliminary testing: The ball may be reused by rotating it to expose a unworn contact surface.
For final measurements: It is recommended to employ a new ball for each test to ensure accuracy and reproducibility.

Extension holder

page icon
  • Use the extension only in particular cases where the ball holder cannot reach the sample.
  • Increasing the ball holder length can negatively affect test results, especially in reciprocating tests. Whenever possible, use the load cell extension block instead.
 
 
notion image
(to reduce the distance towards the sample)

Universal Ball holder Overivew


 
notion image

1. Test Ball or Pin

Rtec balls catalog


Upper Sample Holder for 1000℃ Chamber
Upper Sample Holder for  1000℃  Chamber
Part no.
0.5 inch ball
SPN030060-55
0.375 inch ball
SPN030060-54
0.25 inch ball
SPN030060-53
3 mm ball
SPN030060-56
6mm diameter cylinder
SPN030041-462
6.3mm diameter cylinder
SPN13108
9.5mm diameter cylinder
SPN030041-463
12.7mm diameter cylinder
SPN030041-464
15mm diameter cylinder
SPN030041-465
  • E52100 Alloy Steel / HRC60
  • 304 SSt / HRC25
  • 440C SSt / HRC58
  • WC Tungsten Carbide / HRC75
  • SiN Silicon Nitride
  • Nonporous Alumina Ceramic balls
  • PTFE
Available Ball size :

  • 1.6mm
  • 3.9mm
  • 6.3mm
  • 9.5mm
  • 12.7mm
 

2 . Nut

3. ER11 Collet

General metric range avalaible: from 1 mm to 7 mm (0.5 mm increments)
Each collet has a clamping range of 0.5 mm
ex: an ER11-3 mm collet can also clamp pins/balls with a 2.5-3.5 mm diameter.

4. Adjusting pin

This pin enables ball position adjustment within the collet.

5. Ball Holder

Holder Specification MSC 81197253
Collet Series
ER11
Shank Type
Straight
Through Coolant
No
Shank Diameter (Inch)
5/8
Minimum Collet Capacity (Decimal Inch)
0.0190
Maximum Collet Capacity (Decimal Inch)
0.2760
Overall Length (Inch)
3-1/2
Overall Length (Decimal Inch)
3.5
Actuator Type
Wrench
Shank Length (Decimal Inch)
3.0300
Collet Nut Diameter (Decimal Inch)
0.6200

6. Extension

page icon
For additional information or to place an order, please contact Rtec Support (contact information provided at the end of this manual).
 

Self-Adjusting Block holder preparation


notion image
page icon
The self-leveling block holder will ensure proper contact during the test.
  1. Firstly ,loosen the 2 tightening screws using /16” Allen key.
  1. Slide in the block sample into the block support
    1. Avalaible ASTM Rtec Block Catalog

      ASTM: D3704, G77, G176, D2714, D3704, D2509
  1. Level the block sufficiently into the holder.
  1. Tighten the securing screws on each side.
 

Block sample Quotation


notion image

Rtec Test Block Size: 0.620 x 0.250 x 0.4

L x l x h in inches
Reference : MM000128-XX
notion image
Dimension in inches
Dimension in inches
 

Installing the Ball Holder


  1. Use four 1/4” button head screws to secure the assembly to the load cell
    and tighten using a 5/32” Allen wrench.
    Then Insert the slip sleeve into the sensor mount.
    1. notion image
  1. Place the ball holder into the slip sleeve.
  1. Mount the Ball Holder onto the suspension in the same manner as onto
    the load cell using a 9/64” Allen wrench.
    1. Montage with suspension
      Montage with suspension
      Montage without suspension
      Montage without suspension
 

Mounting the Argon Sensor on MFT-2000


  • Install the load cell on the fast-exchange attachment by fastening the 4 captive screws using a 5/32" Allen wrench.
notion image
page icon
  • The narrow side of the fast exchange plate’s should point to the left of the front load cell as this side will fit into the back of the sliding support.
  • The front of the load cell is the face showing the Rtec logo and the unit calibration sticker.

(Optional) With Extension block:

page icon
(Optional) You can also use an extension block to reduce the distance between the load cell and the lower setup.
2" (left) and 4" extensions (right)
2" (left) and 4" extensions (right)

  1. Mount the block extension on the exchange plate with 4 4 x 10-32 x 1.250” long screws using 5/32 Allen wrench.
    1. notion image
  1. Then the adaptor plate mounted on the extension block with 4 x 10-32 x .625” long screws using 5/32 Allen wrench.
    1. notion image
  1. Install the load cell on the fast-exchange attachment by fastening the 4 captive screws using a 5/32" Allen wrench.
    1. notion image
page icon
  • The narrow side of the fast exchange plate’s should point to the left of the front load cell as this side will fit into the back of the sliding support.
  • The front of the load cell is the face showing the Rtec logo and the unit calibration sticker.
 
 
 

Installing the Argon (MFT-2000)


notion image
notion image
  1. Slide in the load cell into the Z stage rack.
    1. page icon
      Make sure the 4 screws above the rack are removed.
      Slide the load cell with its front facing you and the connector on the right.
  1. Fasten the 4 securing screws by hands.
    1. notion image
  1. Connect the ribbon cable. The connector only fit one way.
 
 
 

Adjust the Y Radius

 
 

Hardware Final Notes


  • Always confirm all screws are hand-tightened and then secured with the appropriate Allen key.
  • Refer to the full user manual for safety precautions and maintenance schedules