- Connect and switch on the Thorlabs microscope.
- Click on HFRR Build Report on the Desktop
or search it on window

- The program is opened on the browser.
- Check first if a lens configuration is already existing.
If there is no existing configuration, you must calibrate it (step detailed further)

- One the tested ball is positioned underneath the microscope, take an image.
Click on LIVE IMAGE to get the live image acquisition

- Crop the Y upper left side, the bottom X right side of the ball.
- If the objectives lens is not calibrated, you must use a calibrated sample as reference, crop it, insert the physical X and Y distance and press calibration. X and Y distance and press calibration.
- Step number 3: Upload the CSV result made with the RTEC Software
- Press PDF Report
Example of PDF Report