Standard Reciprocating Drive Motor #E -SPN04325
Standard Reciprocating Drive Motor #E -SPN04325

Standard Reciprocating Drive Motor #E -SPN04325

Module Installation


Required Tools and Components


Components

  • Linear Reciprocating Drive
  • Direct Rotary Drive
  • Universal Sample Holder
  • Sample
  • Optional Components: LVDT Attachment

Screws and Hardware


  • (7x) Direct Rotary Drive Screws – (310280-05)
  • (1x) Alignment Pin
  • (2x) Reciprocating Drive Screws (310280-12)
  • Allen wrenches: 5/64", 9/64”
 
 

Direct Rotary Drive Installation Step


notion image
page icon
Please skip this step if your drive is already installed onto the XY stage.
As shown above, the drive is installed on the stage.
Additional animation instructions
Additional animation instructions
notion image
 
  1. Route the drive cable through the X Y stage.
  1. Position and insert the motor drive through the stage.
  1. Orient the drive so the green sensor port faces the right side.
notion image
 
notion image
 
Secure the drive with 7 x SHCS 8-32 X .625" long screws  
(310-280-05 / BM310280-09)
notion image
Confirm that the alignment pin is seated correctly.
Confirm that the alignment pin is seated correctly.
Connect the 2 cables on the slot on the right, behind the frame (the Motor Power Chord and the Encoder Chord).
notion image
⚠️
Always power off the instrument before connecting cables or installing any
load cell or accessory.

Hardware Final Notes


  • Always confirm all screws are hand-tightened and then secured with the appropriate Allen key.
  • Refer to the full user manual for safety precautions and maintenance schedules